What checks shall be carried out, while calibrating DT apparatus?

While calibrating DT apparatus, following checks shall be performed.

1.) Number of strokes per minute (Limit: 29-32 cycles/min)
2.) Temperature by probe & standard thermometer (Limit:  37 ± 1 OC).


3). Distance travelled by basket (Limit: 53 -57mm)